Description
Temperature/RH Loggers
These temperature and relative humidity data loggers are housed in a robust, waterproof (IP68-rated) case, which is designed for use in harsh applications like cylinder curing rooms. They are available with an integral, coated RH sensor (H-2736) or with a temperature and relative humidity probe with a 1.5m cable length (H-2732). The relative humidity probe features a coated RH sensor that shows good resistance to moisture and condensation, ensuring measurement reliability. These data loggers have a high reading resolution and accuracy and provide fast off-load speed and a low battery monitor. Data is downloaded to a computer for viewing, reports and archiving. Data stored on the logger will be retained after a battery is replaced. See tables below for specific logger specifications. Require H-2736.SW software and a download cable for operation. Dimensions: 1.75″ x 2.25″ x 3.15″ (34 x 57 x 80mm)
Specifications
Reading Capacity | 32,000 |
Memory Type | Non-volatile 64K |
Reading Types | Actual, Min, Max |
Delayed Start | Relative / Absolute (up to 45 days) |
Trigger Start | Magnetic Switch (H-2736.3) |
Alarms | 2, fully programmable |
Stop Options | When full, After “n” Readings, Never overwrite oldest data |
Operating Range | -40°F to +185°F -40°C to +85°C |
Battery | User-replaceable Lithium |
Temperature Specifications
Sensor | 10K NTC Thermistor |
Range | -13 to 185°F (-25 to 85°C) |
Response Time | 25 mins to 90% FSD in moving air |
Accuracy | 0.01°C or better |
Relative Humidity Specifications
Sensor Type | Capacitive |
Reading Range | 0 to 100% RH |
Accuracy | ±3.0% at 77°F (25°C) |
Reading Resolution | Better than 0.3% RH |
Sensor Location | Externally mounted |
Response Time | 10 seconds to 90% |